Measuring water flow to generating units at the 1,520-MW Manic-5 powerhouse, with the goal of maximizing unit efficiency, is a monumental task but the long-term benefits could help Hydro-Quebec customers and establish industry-wide benchmarks.
By Bertrand Reeb, Guillaume Dubois, Gilles Proulx, Ion Candel and Cornel Ioana
In 2013, personnel with Canadian provincial utility Hydro-Quebec
performed penstock intake flow comparison tests to measure
characteristics of the flow to its turbines at the Manic-5 powerhouse,
part of the Daniel-Johnson hydroelectric facility, constructed between
1959 and 1970. Daniel-Johnson Dam is a multiple-arch buttress dam on the
Manicouagan River that creates the Manicouagan Reservoir, 133 miles
north of Baie-Comeau in Quebec, Canada.
The facility produces hydroelectric power and supplies water to the
Manic-5 and Manic-5-PA powerhouses that have a combined capacity of
2,596 MW. The dam is 702 feet tall by 4,311 feet in length making it the
largest dam of its type in the world and one of the company's largest
hydroelectric plants. Information from the tests allow plant engineers to achieve optimal performance from Manic-5's eight 190-MW Francis turbines.
Measurement objectives
In their efforts to keep the total cost of the measurement tests as
low as possible, engineers carefully considered flow measurement
methods. Engineers consider the Manic-5 design intakes favorable and
decided to use two types of measurements within the intake: the current
meter (CM) and the acoustic scintillation flow meter (ASFM) methods. Flow entering each penstock intake passes through two measurement
systems after moving beyond the trashracks apparatus, but prior to
reaching turbine-generators.
Hydro-Quebec's CM method uses trolleys that support a single row of meters.1
The meters move vertically, allowing velocity sample profiles over the
entire height of the measurement section. This set-up reduces the
installation time because it can be used for flow measurement for as
many as four units at this location (during the test, only one of the
four units was operated). Even though flow measurement is traditionally
considered difficult to perform within intakes, especially in a short
converging intake, the measurement is proving accurate in many
situations such as shown in Kootenay Canal comparative tests2 and
showing many advantages over penstock measurement.2
Within the penstock and aft of the CM measurement system, engineers
used an ASFM to measure discharge as part of a three-year partnership
between Electricite de France's General Technical Division (EDF DTG),
Hydro-Quebec and the manufacturer of the ASFM, ASL AQFLow (the "SMASH"
project).3
The main focus by partnership members studied ASFM in more detail
and, depending on their findings, would possibly extend ASFM's range of
usage. As a matter of fact, the ASFM measurements from the Manic-5 study
now serve as benchmarks for normal measurement conditions.
The power plant
The power plant has eight 190-MW Francis turbines operating with a
net head of 141.8 m (465 feet). Two sets of conduits each provide water
to four units and each conduit has two rectangular bays, which converge
into one circular section downstream from the head gates. The
measurement sections are 10.97 m by 6.1 m (36 feet by 20 feet). The CM
measuring section was located about 10 m from the trash racks in the
stoplog slots, while the ASFM measuring section was 3.4 m farther
downstream in the gate slots (see Figure 1). With only one unit in
operation, the average velocity was in the range of 1 to 2.5 m/sec,
which falls within the usual range of both the CM and ASFM.
The measurement conditions were considered good for both CM and ASFM.
In effect, the smooth transition between the trashracks and the
measurement sections is likely to produce smooth velocity profiles. The
only concern were the trashracks cross members, which can generate large
wakes. This has been taken into account in the selection of the number
of ASFM measurement levels. The flow angle was expected to be near
horizontal.
ASFM
ASFM uses a technique called acoustic scintillation drift to measure the flow velocity perpendicular to a number of acoustic paths established across the intake to the turbine. Short (16 µsec) pulses of high-frequency sound (in the order of 307 kHz) are sent from transmitting arrays on one side to receiving arrays on the other, at a rate of about 250 pings/sec.5
ASFM uses a technique called acoustic scintillation drift to measure the flow velocity perpendicular to a number of acoustic paths established across the intake to the turbine. Short (16 µsec) pulses of high-frequency sound (in the order of 307 kHz) are sent from transmitting arrays on one side to receiving arrays on the other, at a rate of about 250 pings/sec.5
Frame supporting the current meters. The frame had 14 current meters installed on the bottom of a simple trolley made of two end plates, profiled rods and steel cables. |
Fluctuations (known as scintillations) in the amplitude of those
acoustic pulses result from turbulence carried along by the flow. ASFM measures scintillations and from them computes the lateral
average (i.e., along the acoustic path) of the velocity perpendicular to
each path.
In its simplest form, two transmitters are placed on one
side of the measurement section and two receivers at the other. The
signal amplitude at the receivers varies randomly as the turbulence
along the propagation paths changes with time and flow.
A steel frame, which is formed by two main vertical beams, supports 21 ASFM transducers. |
If the two paths are sufficiently close (Δx), the turbulence remains
embedded in the flow, and the pattern of these amplitude variations at
the downstream receiver will be nearly identical to that at the upstream
receiver, except for a time delay, Δt. This time delay corresponds to
the peak in the time-lagged cross-correlation function calculated for
Signal 1 and Signal 2. The mean velocity perpendicular to the acoustic
paths is then Δx/Δt. Using three transmitters and three receivers at
each measurement level allows both the magnitude and inclination of the
velocity to be measured. ASFM computes the flow through each bay of the
intake by integrating the horizontal component of the velocity over the
cross-sectional area of the intake. In a multi-bay intake, the flows
through each bay are summed to compute the total flow.
For the measurement, 21 pairs of transducers were installed on a
steel frame. The number of transducers was chosen in order to resolve
the possible oscillation of the velocity profile due to wake of the
trash racks main cross members. All transducers were set horizontally
because the expected flow angle was horizontal, as it was later
confirmed by the measurement.
This is a close up of horizontally installed ASFM transducers on a steel frame used for flow measurement. |
In addition to the normal ASFM, a high-speed data acquisition system
was used to record the raw acoustic signal for each element. In total,
12 signals were simultaneously recorded at a rate of 3.3 mega samples
per second) for each 7-minute run. Those data were used after the tests
for an alternate post-processing analysis.
CM measurement set up
CM measurement was done with 14 meters installed on the bottom of a
simple trolley made of two endplates, profiled rods and steel cables
(see photos on pages 72 and 73). The CMs were equipped with Type A self-compensating propeller that
compensates for up to 45-degree flow angle. The profiled rods are the
same that are used for calibration of the CMs. Steel wheels help guide
the trolley in the gate slot both laterally and longitudinally. The
frame was moved by means of two hoists controlled by variable-frequency
power drives, which allowed selection of the travel velocity. The two
hoists were synchronized by means of encoders. Two displacement
transducers measured the elevation of each end of the trolley.
The trolley velocity was set to 50 mm/sec except at the top and
bottom of the measurement section, where it was set to 10 mm/sec for
better defining the rapidly changing velocity profile. Total measurement
time was 8 minutes for each run. The trolley velocity for the profiling
method represents about 2% of the average velocity at the lowest flow.
The data acquisition software developed by Hydro- Quebec allows
recording the instantaneous rotational velocity of each current meter,
(i.e., it records the time stamp of each revolution). Once the
rotational velocities are recorded, it is relatively easy to calculate
the mean value for different time intervals.
Results
Due to the presence of trashracks and a rigid and vibration-free
support frame for the ASFM transducers, the conditions for ASFM testing
were very close to ideal, meaning that all flow velocities were used in
computing the flow (see Figure 2 on page 71). The velocity profile
measured by ASFM shows small oscillations that are related to the
trashracks cross members.
CM results show that the velocity profiles produced are similar to
the ones from ASFM, with ASFM results slightly higher. The small
difference (higher velocities in the bottom part of the measurement
section) can result from the ASFM measurement section being farther
downstream than the CM section, as the velocity profile tends to develop
or flatten as the turbulence is mixing the different layers of the
flow. A part of the difference can also be due to two other factors: the
acquisition times for the CMs and ASFM are not concurrent and the sizes
of the measurement sections of the two methods are marginally
different.
The three-dimensional velocity profile (also Figure 2) from the CM
measurement shows some asymmetry, especially in the bottom part of the
section. It is likely due to the asymmetry of the intake upstream of the
trashracks or the presence of some debris there. This asymmetry of the
velocity profile in the bottom part of the measurement section can be
the source of difference between the CM and ASFM methods.
Flow measurement comparison
The comparison of the flow measured by CM and ASFM (see Figure 6) includes the results for four units and 40 runs. Overall, the difference between the two methods is 1.4 %, with the ASFM flow being higher. The standard deviation is ± 0.8 % and includes the deviation of both methods. The random uncertainty of the regression line is ± 0.3 %, which means that there exists a statistical difference of 1.1% between the two methods. There is no significant variation of the difference between the two methods as a function of the flow. Both methods have an expected uncertainty of ± 1% to ± 1.5%. This means each method's uncertainty band overlaps. Thus, project engineers consider the agreement between the two methods as "good."
Reprocessing ASFM data
Project personnel compared the standard time series computed by ASFM and the recomputed time series based on the recorded raw signals. They used a new algorithm to determine the amplitude of the acoustic pulses. This algorithm has already shown to improve the results of ASFM in case of very low turbulence3, which is not within the normal range of usage of the method.
This new algorithm is not expected to have a significant effect on
the calculated flow under normal stipulated conditions of usage of ASFM.
Figures 3, 4 and 5 (see pages 72, 73 and 74) show a very high
resemblance between the two time series, which leads to close values for
the total flow. The quality factor of the measurement, for each
individual level5 was above 0.9, thus a high degree of confidence in the
ASFM results.
From the eight runs reprocessed with the new algorithm, the
difference from the original calculation is 0.35%, which is low. As
expected, the difference for the present tests is not large, unlike what
was found for low turbulence level tests. The new algorithm will be
tested more thoroughly in future ASFM comparison tests as Hydro-Quebec,
EDF and ASL AQFlow continue to collaborate on this matter.
Comparison of efficiency
Figure 7 (see page 79) shows the efficiency of Unit 4 (this unit
represents an average one from the point of view of flow comparison
between the CM and ASFM). The efficiency measured with the CM and ASFM
in 2013, the thermodynamic method (Unit 8) measured in 1992 and the
expected efficiency obtained from the model test step up are plotted as a
function of the turbine output. At peak efficiency, all four curves are
within 1.4%. Both the CM and ASFM intake measurement methods can be
considered to have a measurement uncertainty of about ±1 to ± 1.5%.
At peak efficiency, the difference between the eight units reaches
0.3 % and will be taken into account for optimal dispatching. There is
also some difference between the powers at peak efficiency, which will
also be taken into account for turbine set up. In addition, for turbine
output toward the maximum values, the difference of efficiency reaches
1.5% between the most and least efficient units. Again, knowing this
fact will help Hydro-Quebec to properly operate its units, because they
are often used during high load periods.
Conclusions
Hydro-Quebec has performed efficiency tests in the intake of one of
its major power plant. The goal of those tests was to measure the
efficiency of all eight units for optimal dispatching. It was also a
good opportunity to perform comparative test between two methods, CM
ASFM.
The results show that the flow measured by the CMs and ASFM are
within 1.4% from each other, which is considered excellent. The results
of the CMs and ASFM also compare well with the previous tests done with
the thermodynamic method. Significant differences exist between each unit at peak efficiency and near maximum power. The results of those tests will help Hydro-Quebec to better operate this plant.
Notes
- Lamy, P., and J. Neron, "A Different Approach in Measuring Individual Turbine Efficiencies in Multiple Unit Power Plants," Proceedings of Water Power XIII, HCI Publications, Kansas City, Mo., 2003.
- Proulx, G., and E. Cloutier, "Hydro-Québec Experience with Discharge Measurement in Short Converging Intake," Proceedings of HydroVision International 2011, PennWell, Tulsa, Okla., 2011.
- Candel, I., Bertrand Reeb, David Lemon and Cornel Ioana, "Electricité de France's Study of the Acoustic Scintillation Flow Meter Results in Expanding its Range and Sensitivity," Proceeding of Hydro2013 Conference, Innsbruck, Austria, 2013.
- ASFM Operations Manual, AQFlow.
- Lemon, D.D., D. R. Topham, and D. Billeness, "Improvements To The Accuracy Of Discharge Measurements By Acoustic Scintillation Resulting From Revisions To Data Processing Procedures," Proceeding of IGHEM 2010 Conference, International Group for Hydraulic Efficiency Measurement, Roorkee, 2010.
- Almquist, C.W., Taylor, J.-W., "Results of Kootenay canal flow comparison test using intake methods." Proceedings of Hydro2010 Conference, Lisbon, Portugal, 2010.
Briefly
Several methods exist to measure absolute water flow rate in hydro
plants, some of which include the current meter, pressure-time (Gibson),
tracer and ultrasonic methods. The newest is the ultrasonic method and
it is being applied more frequently at hydro plants because it allows
for continuous monitoring of the flow rate.
However, the basic flow
measurement methods used for efficiency tests of hydraulic machines are
the current meter and Gibson methods. The current meter method has frequently been applied at hydroelectric
facilities. Another possibility for continuous measurement of the
discharge is the non-absolute Winter-Kennedy method.
http://www.renewableenergyworld.com/articles/print/hydro/volume-34/issue-6/articles/flow-tests-lead-to-increased-efficiencies-and-benchmark-measurements.html
No comments:
Post a Comment